HOLOGIC, INC SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 3D Back to Search Results
Model Number SDM-SYS-6000-3D
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/23/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the c-arm auto move down.No injury reported.A field engineer was dispatched to the site and determined the footswitch needed to be replaced and the cable in pmc needed to be reseated.Once this was completed the system was working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 3D
Common Device Name
MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
MDR Report Key9424180
Report Number1220984-2019-00138
Device Sequence Number533878
Product Code OTE
UDI-Device Identifier15420045505636
UDI-Public(01)15420045505636
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
P080003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,user faci
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSDM-SYS-6000-3D
Device Catalogue NumberSDM-SYS-6000-3D
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2019
Initial Report FDA Received Date12/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-