| Model Number |
2AF284 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Cardiac Perforation (2513); No Code Available (3191)
|
| Date of Event |
11/27/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: flexcath advance steerable sheath.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during a cryo ablation procedure, the mapping catheter extended out and punctured the lower lobe of the patient's right lung, which began to fill up with blood.The case was aborted with the patient under general anesthesia.Blood was backed up into the endotracheal tube.A bronchoscopy was performed to remove the blood from the patient's lung.It was noted the right inferior pulmonary vein (ripv) was possibly perforated as well.The patient's hospitalization was extended.The patient was stable at the time of the report, but their recovery time was unknown.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the data files and balloon catheter 2af284 with lot number 14114, were returned and analyzed.The data files showed that at least ten applications were performed with balloon catheter, 2af284 with lot number 14114, without any issue on the date of the event.Visual inspection of the balloon catheter showed the mapping catheter was inserted into the catheter and it was intact with no apparent issues.Smart chip verification indicated the catheter was used for 10 injections.The catheter passed the performance test and electrical integrity as per specification.Clinical issues were encountered during the case.In conclusion, the clinical issues were unable to be confirmed via data analysis or product analysis.The balloon catheter passed the returned product inspection as per specification.There is no indication of relation of adverse event to the performance and malfunction of the product.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Incoming information indicates the patient recovered.
|
| |
|