• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION TMJ SYSTEM RIGHT STANDARD MANDIBULAR COMPONENT; JOINT, TEMPOROMANDIBULAR, IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION TMJ SYSTEM RIGHT STANDARD MANDIBULAR COMPONENT; JOINT, TEMPOROMANDIBULAR, IMPLANT Back to Search Results
Model Number N/A
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Information (3190)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2019-00542.Explantation date is planned for (b)(6) 2020.Medical product: tmj system right fossa component, small, part# 24-6562, lot# 630060b, tmj system right standard mandibular component, part# 24-6550, lot# 845440a, unknown screws, part# unk, lot# unk.The user facility is foreign; therefore a facility medwatch report will not be available.Report source: (b)(6).
 
Event or Problem Description
It was reported a revision is planned due to malocclusion.The explanted devices will be replaced by a custom implant.No additional patient consequences have been reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The complaint is considered confirmed as a revision surgery was reported.No product was returned; therefore, no functional tests or inspections could be performed.No x-rays, scans, pictures, or physician's reports were provided.The non-conformance database (tipqa) was reviewed for the mandible component; no non-conformances were found.There are no indications of manufacturing defects.For all non-custom tmj mandibular implants in the previous one year (from the notification date) regarding malocclusion, there is a complaint rate of (b)(4) , which is no greater than the occurrence listed in the afmea.The most likely underlying cause of the complaint is improper placement of the implant(s), though it was not clear from the information received if it was the position of the fossa, mandible, or both implants that led to the malocclusion.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.The following fields were updated: b4 date of this report, b5 describe event or problem, d4 expiration date, d10 device availability, g4 date received by manufacturer, g7 type of report, h2 follow up type, h3 device evaluated by manufacturer, h6 method code, h6 results code, h6 conclusions code, and h10 additional narratives/data.
 
Event or Problem Description
This follow-up report is being submitted to relay additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TMJ SYSTEM RIGHT STANDARD MANDIBULAR COMPONENT
Common Device Name
JOINT, TEMPOROMANDIBULAR, IMPLANT
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key9428388
Report Number0001032347-2019-00543
Device Sequence Number19260306
Product Code LZD
UDI-Device Identifier00841036060343
UDI-Public00841036060343
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P020016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/20/2023
Device Model NumberN/A
Device Catalogue Number24-6550
Device Lot Number845440A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/11/2019
Supplement Date Received by Manufacturer04/24/2020
Initial Report FDA Received Date12/06/2019
Supplement Report FDA Received Date05/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Required Intervention;
-
-