|
Zimmer biomet complaint (b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2019-00533, 0001032347-2019-00534, 0001032347-2019-00536.Medical products: 2.0mm system twist drill with j notch, part# 01-9198, lot# 753330, 2.0mm system twist drill with j notch, part# 01-9198, lot# unknown, 2.0mm system twist drill with j notch, part# 01-9203, lot# unknown, 2.0mm system twist drill with j notch, part# 01-9203, lot# unknown.The user facility is foreign; therefore a facility medwatch report will not be available.Report source: (b)(6).
|
|
This follow-up report is being submitted to relay additional information.The complaint is confirmed.On visual inspection, the drill was noted to have a transverse fracture through the flutes near the drill neck.The cert was reviewed for 01-9203 serial number (b)(6) and there were no non-conformances found.There are no indications of manufacturing defects.This is the only complaint for 01-9203 serial number (b)(6)regarding the drill fracturing.For this part (01-9203) and the previous three years (from the notification date) regarding the fracturing of this drill, there is a complaint rate of (b)(4) which is no greater than the occurrence listed in the application fmea.The most likely underlying cause of the complaint is the drill experienced force in excess of what it was designed to encounter; fractures at this point of the drill are typical of excessive force fractures.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|