| Catalog Number |
CDS0501 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
Thrombosis (2100)
|
| Date of Event |
11/13/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
This is being filed to report the foreign material and clots.It was reported that this was a mitraclip procedure to treat mixed mitral regurgitation (mr) with a grade of 3-4.The clip delivery system (cds) was advanced into the patient anatomy when an object was noted on the clip arms and thrombus was suspected.The cds was removed and it was noted the clip arms were entangled with a silk like object with some clots.The object was rinsed and the physician tried to break it but was unsuccessful.The object was sent to testing within the hospital.The procedure was continued with the one clip implanted, reducing mr to 1+.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The complaint device was returned for evaluation.In this incident the return device analysis was unable to confirm the reported device contamination.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar incidents reported from this lot.The reported patient effect of emboli (thrombosis), as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures.All the available information was investigated and the definitive cause for the reported device contamination and thrombosis could not be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Search Alerts/Recalls
|