MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP SYSTEM CLIP DELIVERY SYSTEM; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP SYSTEM CLIP DELIVERY SYSTEM; VALVE REPAIR Back to Search Results
Catalog Number CDS0501
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code Thrombosis (2100)
Date of Event 11/13/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
This is being filed to report the foreign material and clots.It was reported that this was a mitraclip procedure to treat mixed mitral regurgitation (mr) with a grade of 3-4.The clip delivery system (cds) was advanced into the patient anatomy when an object was noted on the clip arms and thrombus was suspected.The cds was removed and it was noted the clip arms were entangled with a silk like object with some clots.The object was rinsed and the physician tried to break it but was unsuccessful.The object was sent to testing within the hospital.The procedure was continued with the one clip implanted, reducing mr to 1+.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Additional Manufacturer Narrative
Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The complaint device was returned for evaluation.In this incident the return device analysis was unable to confirm the reported device contamination.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar incidents reported from this lot.The reported patient effect of emboli (thrombosis), as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures.All the available information was investigated and the definitive cause for the reported device contamination and thrombosis could not be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM CLIP DELIVERY SYSTEM
Common Device Name
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9429133
Report Number2024168-2019-14237
Device Sequence Number469377
Product Code NKM
UDI-Device Identifier08717648216800
UDI-Public08717648216800
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/16/2020
Device Catalogue NumberCDS0501
Device Lot Number90416U138
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2019
Initial Date Received by Manufacturer 11/14/2019
Supplement Date Received by Manufacturer01/03/2020
Initial Report FDA Received Date12/06/2019
Supplement Report FDA Received Date01/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Weight47
-
-