Catalog Number 2C8541 |
Device Problems
Fluid/Blood Leak (1250); Component Missing (2306)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/11/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a clearlink duo-vent continu-flo solution set was leaking.It was further reported that the medication was coming out of one of the ports because the ¿valve¿ (of the clearlink port) was missing.This issue was identified during priming.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The actual sample was received for evaluation.A visual inspection was performed, and it was noted that there was leak at the y-site as the second y-site was not assembled correctly.The reported problem was verified.The cause of the condition was due to manufacturing related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|