| Model Number |
2000 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
07/30/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that detachment of device occurred.The target lesion was located in the coronary artery.A floppy 182cm j choice guidewire was used for treatment.However, during procedure, entrapment occurred when removing the device.The guidewire tip fracture remained inside the patient's body.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device was returned for analysis.The guidewire returned inside of a protective hoop and together with its original opened cardboard box.The guidewire was easily removed from the protective hoop.The guidewire was inspected and the distal tip was found stretched.No more visual damages were found in the device.Microscopic inspection revealed that the distal tip of the guidewire was inspected and it was possible to observe that the spring was stretched.The core wire was detached from the distal end solder ball; the solder ball remained attached to the spring tip.Dimensional inspection revealed that the overall length and distal tip outer diameter (od) was not within specifications due to the stretched spring tip.The middle section od and the proximal section od were within specifications.
|
| |
|
Event or Problem Description
|
|
It was reported that detachment of device occurred.The target lesion was located in the coronary artery.A floppy 182cm j choice guidewire was used for treatment.However, during procedure, entrapment occurred when removing the device.The guidewire tip fracture remained inside the patient's body.
|
| |
|
Search Alerts/Recalls
|