| Model Number |
CAT02298 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
11/15/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a potential for foreign material being left in the joint.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: black foreign material in lesion probable root cause: design - instrument tip not designed for insertion into cannula - improper raw material selection - assembly design not strong enough to withstand insertion effort manufacturing - incorrect raw materials used - instrument not manufactured to specification - instrument not assembled properly application - excessive force - user attempts to insert into cannula which already supports access of other instruments to the joint space - user attempts to insert device without use of a cannula the reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a potential for foreign material being left in the joint.
|
| |
|
Search Alerts/Recalls
|