It was reported the patient suffered a traumatic injury due to a car accident and developed hydrocephalus.On december 4, the patient was implanted with an adjustable anti-siphon shunt.Intraoperatively, the ventricle end catheter and abdominal catheter were connected normally.After pressing the valve reservoir, a small amount of cerebrospinal fluid was found at the end of the valve.The valve was replaced with a new one at the doctor's request.The patient's status at the time of the report was alive-no injury.
|
|
The returned valve was patent.It met the requirements for valve flux, siphon control and preimplantation testing.The valve did not meet the requirements for leak testing due to a tear on the delta chamber.The valve did not meet the requirements for reflux and pressure flow testing.This tear was not evidently visible during visual inspection but was detected during leakage test as air bubbles were found near the delta chamber surface.It is unknown how or when this damage occurred.The instructions for use cautions, ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting or crushing of components.Such damage may lead to loss of shunt integrity¿.¿ dhr review performed on lot # e45625 did not indicate any finding correlated to the complaint.All valves are 100% tested at the time of manufacture.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|