Model Number 42866
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported the patient suffered a traumatic injury due to a car accident and developed hydrocephalus.On december 4, the patient was implanted with an adjustable anti-siphon shunt.Intraoperatively, the ventricle end catheter and abdominal catheter were connected normally.After pressing the valve reservoir, a small amount of cerebrospinal fluid was found at the end of the valve.The valve was replaced with a new one at the doctor's request.The patient's status at the time of the report was alive-no injury.
 
Additional Manufacturer Narrative
The returned valve was patent.It met the requirements for valve flux, siphon control and preimplantation testing.The valve did not meet the requirements for leak testing due to a tear on the delta chamber.The valve did not meet the requirements for reflux and pressure flow testing.This tear was not evidently visible during visual inspection but was detected during leakage test as air bubbles were found near the delta chamber surface.It is unknown how or when this damage occurred.The instructions for use cautions, ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting or crushing of components.Such damage may lead to loss of shunt integrity¿.¿ dhr review performed on lot # e45625 did not indicate any finding correlated to the complaint.All valves are 100% tested at the time of manufacture.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA II VALVE, REGULAR
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key9439419
Report Number2021898-2019-00434
Device Sequence Number2474165
Product Code JXG
UDI-Device Identifier00643169975835
UDI-Public00643169975835
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K042465
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model Number42866
Device Catalogue Number42866
Device Lot NumberE45625
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2019
Supplement Date Received by Manufacturer03/19/2020
Initial Report FDA Received Date12/09/2019
Supplement Report FDA Received Date03/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-