• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYSMEX RA CO. LTD. CA-620; AUTOMATED BLOOD COAGULATION ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYSMEX RA CO. LTD. CA-620; AUTOMATED BLOOD COAGULATION ANALYZER Back to Search Results
Model Number CA-620
Medical Device Problem Codes Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/05/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The sysmex ca-600 series instructions for use (ifu), chapter 12 - troubleshooting, section 12.1 - introduction, informs the user: "when an error occurs in a sample, the analysis data relative to the error is displayed "***.Repeat analysis is required in the case of either an instrument error or an analysis error (shown as "***.In the stored data and on print out)." additionally, analysis data error, informs the user of error codes and messages that may accompany a result and provides probable cause for the error as well as actions / steps to undertake to resolve the issue.For the "clotting method", the "no coagulation flag" err[32], the ifu indicates: "coagulation reaction can not be detected".The ifu further states as an action/step: "make comprehensive judgment, taking various aspects into consideration, including whether or not any error is involved in samples (storage status, blood collecting procedure, etc.) and reagents (storage status)." the analyzer performed as designed by alerting the operator that results needed verification prior to reporting.The operator failed to follow the troubleshooting steps for the "no coagulation" flag and did not verify the results prior to release.
 
Event or Problem Description
A sample was analyzed and generated asterisks with a "no coagulation" flag instead of a numeric protime (pt) result.The user reported the pt result as "no coagulation" to the physician.The patient was administered 5mg vitamin k (mephyton).No harm to the patient due to the administration of vitamin k was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CA-620
Common Device Name
AUTOMATED BLOOD COAGULATION ANALYZER
Manufacturer (Section D)
SYSMEX RA CO. LTD.
1850-3 hirookanomura
shiojiri, 399-0 702
JA  399-0702
MDR Report Key9440312
Report Number1000515253-2019-00020
Device Sequence Number3999317
Product Code GKP
UDI-Device Identifier04987562425686
UDI-Public0104987562425686
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K031377
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberCA-620
Device Catalogue NumberBR142626
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2019
Initial Report FDA Received Date12/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age85 YR
-
-