| Model Number |
989605455171 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The service engineer tested the system and was able to confirm the reported problem.A visual inspection revealed deposits inside the transducer connector which caused a poor connection between the system and the transducer.After carefully removing the deposits and cleaning the contacts of the connector, the system and transducer passed all tests and resumed normal operation.Both the system and transducer remain in use at the customer site with no return anticipated; therefore, no further failure analysis can be performed.
|
| |
|
Event or Problem Description
|
|
A customer reported encountering an incident where their epiq ultrasound system would not recognize the x8-2t tee model transducer during an elective aortic valve replacement.The failure occurred while the transducer was inserted in the patient, but prior to the bypass surgery.After rebooting the system twice, the transducer became fully functional and the procedure was successfully completed with no adverse effect to the patient.
|
| |
|
Search Alerts/Recalls
|