• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Device Problems Mechanical Problem (1384); Appropriate Term/Code Not Available (3191)
Patient Problem No Information (3190)
Event Date 11/06/2019
Event Type  malfunction  
Event Description
The recipient reportedly experienced intermittent lock.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
The external visual inspection revealed that the array was severed prior to receipt.In addition, the electrode ground ring sleeve was dislodged.These anomalies are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.The electrode condition prevented some electrical tests from being performed.The device passed some of the electrical tests performed.This is an interim report.
 
Manufacturer Narrative
Additional information obtained reportedly indicates that the reason for explant was due to loss of lock.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the array was severed prior to receipt.In addition, the electrode ground ring sleeve was dislodged.These anomalies are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented some electrical tests from being performed.The device passed the electrical and mechanical tests performed.The scanning electron microscopy analysis revealed cracked conductive epoxy at the feedthru pin connection on some pins.This device malfunctioned due to a crack in the conductive epoxy at the feed thru pin-to-substrate connection.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key9440885
MDR Text Key176308062
Report Number3006556115-2019-00729
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup,Followup
Report Date 11/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/01/2006
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/09/2019
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received01/07/2020
02/13/2020
04/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age47 YR
-
-