Model Number 18750S
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior to use, the device could not be deflated completely.There was no patient involvement.
 
Additional Manufacturer Narrative
Additional information: d10, g4, h3, h6 h3 evaluation summary: one sample of device was received for evaluation.The stylet was received as well, a visual inspection was performed and it was observed all the components are assembled properly at the tubes according to drawing.An inflation/deflation test was performed using a syringe air was applied to the cuff and no restriction was observed at the time of inflation neither deflation.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
MMJ SA DE CV(USD)
ave henequen no 1181 desarrol
ciudad juarez,mx 32590
MX  32590
MDR Report Key9441139
Report Number2936999-2019-01005
Device Sequence Number11812730
Product Code BTR
UDI-Device Identifier20884521620145
UDI-Public20884521620145
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K090352
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/08/2023
Device Model Number18750S
Device Catalogue Number18750S
Device Lot Number18A0176JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2020
Initial Date Received by Manufacturer 11/22/2019
Supplement Date Received by Manufacturer02/17/2020
Initial Report FDA Received Date12/09/2019
Supplement Report FDA Received Date03/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-