According to the reporter, during dialysis treatment, there was leakage as there was found to be a pinhole in the silicone extension of the venous lumen under the clamp.This was evident as the catheter was under higher pressure from the dialyzer.It was stated that the pinhole was not noticed during catheter insertion after flushing was done.There was nothing unusual observed on the device prior to use.The catheter was not repaired.The device was replaced in order to resolve the issue.There was no patient injury.
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted: the catheter appeared intact.Pmv performed functional testing which included submerging the catheter into a water bath.The end was clamped, and a syringe was used to inject air to observe leakage.No air bubbles were present.Both extensions were tested with acceptable results.A test guide wire was inserted into the red and blue luer adapters, it passed through with no occlusion.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|