• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SCREW BIOSURE REGENESORB 6MM X 20MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SCREW BIOSURE REGENESORB 6MM X 20MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72204391
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/14/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that during an acl procedure the screw broke.It is unknown if there was a delay or back-up device available.No patient injuries reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported 6x20mm biosure regenesorb screw, intended for use in treatment has not been returned for evaluation.Without the reported product a visual evaluation cannot be performed and the customers complaint cannot be confirmed.From the information provided, the screw broke during use.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: application of excessive force during use.The instruction for use outline precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.A review of the complaint and manufacturing records was performed and there were no indications that would suggest that the device did not meet product specifications upon release into distribution.Further investigation is not warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW BIOSURE REGENESORB 6MM X 20MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key9443028
Report Number1219602-2019-01536
Device Sequence Number2093914
Product Code MAI
UDI-Device Identifier00885554036671
UDI-Public00885554036671
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K142948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/29/2022
Device Model Number72204391
Device Catalogue Number72204391
Device Lot Number50784891
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/14/2019
Supplement Date Received by Manufacturer01/09/2020
Initial Report FDA Received Date12/10/2019
Supplement Report FDA Received Date02/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-