| Model Number |
G48025 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
510(k)#: k163468 investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
The stent was put down into the patient, but would not deploy.Procedure was completed using another stent."as per complaint form"; staff advised the stent was in position over the guidewire and when they removed the red indicator tab to squeeze the trigger the stent failed to deply at all.Another stent was used to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
510(k)#: k163468.Lab evaluation completed on 05-dec-2019 that confirmed the flexor did not kink/break.This file was initially assessed as "flexor kinked/broken" which is a malfunction precedence.This file has been re-assessed based on the lab evaluation details and no longer meets the fda mdr reporting criteria of a malfunction report.
|
| |
|
Event or Problem Description
|
|
This report is being submitted to cancel the initial report submitted based on the receipt of additional information.
|
| |
|