| Model Number |
BONE CUTTER,5.5MM X 13CM |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
11/18/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the arthrex bone cutter ar-8550bc was used during an acromioplasty surgery.During the milling mode the chip of the device broke off.It was not possible to remove the broken fragment.It will remain inside the patient.The surgery was completed successfully with the complaint device.No further information received at the moment.Arthrex gmbh will ask the customer what is meant by the chip of the device.Additional information obtained 03-dec-2019: according to the customer there were metal shavings seen and the will remain inside the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Visual evaluation revealed the presence of horizontal grooves opposite of the cutting window across the ar-8550bc inner tube.Corresponding grooves were identified within the inner diameter of the outer tube window, and consistent with a site of metal debris production.This event is consistent with user applied mechanical force via prying/leveraging, as further evidenced by the presence of scuffing along the shrink tubing at the proximal end of the device.
|
| |
|