Model Number CI-1400-02H |
Device Problems
Mechanical Problem (1384); Defective Device (2588)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 11/11/2019 |
Event Type
malfunction
|
Event Description
|
The recipient reportedly experienced open electrodes.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the array was severed prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed broken electrode wires at the fantail region.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed the electrical and mechanical tests performed.The scanning electron microscopy analysis of the electrode revealed evidence of fatigue breaks.The photographic inspection and scanning electron microscopy analysis revealed all broken electrode wires in the fantail region.It is believed that these breaks caused the multiple electrodes with open circuits.A corrective action was implemented.This is the final.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|