Additional Manufacturer Narrative
|
|
(b)(4).Report source: (b)(6).The investigation is in process.Once the investigation has been completed, a follow-up/final report will be submitted.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
It was reported the patient presented back at the hospital with pain and complaining of impingement after insertion of a tibial nail.Surgeon suspects that nail length is causing the problem.No further information has been made available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further information has been made available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.D4: (b)(4).Reported event was unable to be confirmed due to limited information received from the customer.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.It was reported that the nail implanted was too long but not changed at the point of surgery.The incorrectly sized nail can contribute to impingement and pain.The root cause of the reported issue is attributed to use error - incorrect length nail was implanted.Zimmer biomet offers a range of sizes to choose from.It is up to the surgeon's expertise to choose the appropriate size of the implant to match the patient's anatomy.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.B3 ¿ unknown day in (b)(6) 2020.D7 ¿ unknown day in may 2020.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that patient underwent nail procedure.Subsequently, the nail was explanted approximately 10 months later and no new metal ware was put in.No further event information is available at the time of this report.
|
| |
|