| Catalog Number |
68-750 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified crease fold and an opening on anterior.Leak test and microscopic analysis were performed which identified a puncture opening and observed void >1 mm in non-textured valve-to shell-bond area.Fill test inspection was performed and the result was no blockage.Based on the device analysis the final assessment was a striated punctured due to surgical damage-like consistent in the use of some surgical tools.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device has been explanted.
|
| |
|