Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: analysis of the returned device identified: underweight, opening on posterior, red particles.A visual and microscopic analysis was performed which identified: sharp opening, crease flat.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a sharp opening on posterior due to an unidentified (tear) opening.The reason for reoperation: rupture.This is a known potential adverse event addressed in the product labeling.
|