• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEODENT - JJGC S.A. CM TITAMAX EX IMPLANT 3.75X17 MM; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. CM TITAMAX EX IMPLANT 3.75X17 MM; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 109.670
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Fistula (1862); Pain (1994); Tissue Damage (2104)
Date of Event 10/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).After the evaluation was noticed that the item received is different from the item reported on guarantee form by the complainant.Therefore, the lot number was not considered.Considering the surgical methodology, it was seen that the dentist has used sequence of drills different than the one recommended for the implant installation.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form and visual conference of the product returned by the dentist.
 
Event or Problem Description
(b)(4)¿ the dentist reported that 3 months and 13 days after the dental implant was installed in patient¿s mouth, its non- osseointegration was observed.Moreover, 25n.Cm of primary stability was achieved, the patient presented bone type ii, bone graft was placed and immediate implant was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM TITAMAX EX IMPLANT 3.75X17 MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key9446582
Report Number3008261720-2019-06444
Device Sequence Number2209740
Product Code DZE
UDI-Device Identifier07898237569178
UDI-Public07898237569178
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K101945
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 12/10/2019
Report Date (Section F) 11/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number109.670
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/10/2019
Event Location Other
Date Report to Manufacturer11/14/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2019
Initial Report FDA Received Date12/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR
-
-