| Model Number |
7D63-21 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.There is no further patient information provided due to privacy issues.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely depressed creatine kinase (ck) on one patient generated on the architect analyzer.The results provided were: auto dilution 1:10 = 477.4iu/l / manual dilution 1:10 =32619.1iu/l.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
Through troubleshooting, the customer replaced the physiological saline in the r1 reagent chamber, which resolved the issue.Return testing was not completed as returns were not available.A review of tickets determined normal complaint activity for lot 07560ui19, and no trends were identified for the product for the complaint issue.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.A review of the architect system operations manual provides multiple probable causes and corrective actions for controls being out of range or depressed concentration.The issue was caused by the saline in the r1 reagent chamber and not the creatine kinase reagent.Based on the investigation no product deficiency was identified for the creatine kinase, lot 07560un19.Correction to d suspect medical device, 4.Exp date from 01jan2020 to 02jan2020.
|
| |
|