BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number 24628
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/29/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported that a balloon rupture occurred.The 75% stenosed target lesion was located in the moderately tortuous cephalic vein.A 6.00mm/ 2.0ccm/ 50cm peripheral cutting balloon was selected for use.During the procedure, at first inflation, it was noted that the balloon ruptured at 6atm for several seconds.The procedure was completed with a different device.There were no patient complications reported.
 
Manufacturer Narrative
E1 - (b)(6).Device evaluated by manufacturer: the device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 1mm distal to the distal end of the proximal markerband.An examination of the balloon material and proximal markerband identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 10 atmospheres as per pcb2cm specification.A visual and microscopic examination observed no damage to the tip or blades.All blades were present and fully bonded to the balloon surface.A visual and tactile examination identified no kinks or damage to the shaft of the device.No other issues were identified during the product analysis.
 
Event Description
It was reported that a balloon rupture occurred.The 75% stenosed target lesion was located in the moderately tortuous cephalic vein.A 6.00mm/ 2.0ccm/ 50cm peripheral cutting balloon was selected for use.During the procedure, at first inflation, it was noted that the balloon ruptured at 6atm for several seconds.The procedure was completed with a different device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2CM PERIPHERAL CUTTING BALLOON
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9447413
MDR Text Key176226333
Report Number2134265-2019-15422
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
PMA/PMN Number
K070951
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 02/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24628
Device Catalogue Number24628
Device Lot Number0023948690
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2019
Date Manufacturer Received01/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
6F 3CM MEDIKIT - INTRODUCER SHEATH; 6F 3CM MEDIKIT - INTRODUCER SHEATH; EVEREST - INFLATION DEVICE; EVEREST - INFLATION DEVICE; FUGA - GUIDEWIRE; FUGA - GUIDEWIRE
-
-