| Model Number |
MMT-945 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Unomedical reference number (b)(4).Event occurred in the united states.It was reported that one of the infusion set's tubing was not connected to the needle.The site of insertion was the patient's abdomen and the blood glucose level was 219 mg/dl at the time of incident.No further information available.
|
| |
|
Event or Problem Description
|
|
On 19-may-2020: follow up information was submitted because a visual inspection and test for static pull was performed of the returned used device (1 used set) and the result of complaint investigation showed that all test results were within specifications.Based on the result: device, method, result and conclusion codes were updated.Unomedical reference number (b)(4).Event occurred in the united states.It was reported that one of the infusion set's tubing was not connected to the needle.The site of insertion was the patient's abdomen and the blood glucose level was 219 mg/dl at the time of incident.No further information available.On 19-may-2020: follow up information was submitted because a visual inspection and test for static pull was performed of the returned used device (1 used set) and the result of complaint investigation showed that all test results were within specifications.Based on the result: device, method, result and conclusion codes were updated.
|
| |
|