Device code(s): appropriate term/code not available (3191) was selected for the alleged device tilt.Device code(s): appropriate term/code not available (3191) was selected for the alleged perforation.Investigation is reopened due to additional information provided.The following allegations have been investigated: vena cava perforation, tilt, stroke, atrial fibrillation, depression.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported stroke, atrial fibrillation, depression are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog and lot numbers are unknown, however, the alleged celect is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
Patient allegedly received an implant on (b)(6) 2011 via unknown access due to post trauma.Patient is alleging tilt and vena cava perforation.The patient further alleges atrial fibrillation, stroke and depression.Per a ct (computed tomography) scan of the abdomen dated (b)(6) 2017, ¿no obvious acute abnormalities.Inferior vena caval filter with apex just below the left renal vein.Filter apex tilted to the left.Tips of 4 of the struts are seen outside the wall of the cava.No obvious hematoma seen.No change in the position of the filter when compared with the prior study.¿.
|