• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; MONOFOCAL IOLS Back to Search Results
Model Number ZA9003
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643); No Code Available (3191)
Date of Event 11/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age/ date of birth: unknown/ not provided.Sex/ gender: unknown/ not provided.If implanted, give date: not applicable, the lens was removed/ replaced within the initial procedure.If explanted, give date: not applicable, the lens was removed.Patient left without a lens.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a za9003 intraocular lens (iol) had to be removed upon insertion, because the patient's posterior capsular tear requiring anterior vitrectomy.The customer indicated that patient's capsular bag has inadequate support for a sulcus fixated iol and three piece lens was inserted but was tilting; therefore, iol was removed successfully.A suture was required for closure and patient was sent to a retinal specialist for further evaluation.No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the product was not returned to the manufacturing site; therefore, the complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no additional complaint was received from this production order.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key9450209
Report Number2648035-2019-01306
Device Sequence Number19593002
Product Code HQL
UDI-Device Identifier05050474529144
UDI-Public(01)05050474529144(17)220911
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/11/2022
Device Model NumberZA9003
Device Catalogue NumberZA90030265
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/18/2019
Supplement Date Received by Manufacturer01/24/2020
10/25/2020
Initial Report FDA Received Date12/11/2019
Supplement Report FDA Received Date02/20/2020
11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-