| Model Number |
ZA9003 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Capsular Bag Tear (2639); Vitrectomy (2643); No Code Available (3191)
|
| Date of Event |
11/08/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/ date of birth: unknown/ not provided.Sex/ gender: unknown/ not provided.If implanted, give date: not applicable, the lens was removed/ replaced within the initial procedure.If explanted, give date: not applicable, the lens was removed.Patient left without a lens.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a za9003 intraocular lens (iol) had to be removed upon insertion, because the patient's posterior capsular tear requiring anterior vitrectomy.The customer indicated that patient's capsular bag has inadequate support for a sulcus fixated iol and three piece lens was inserted but was tilting; therefore, iol was removed successfully.A suture was required for closure and patient was sent to a retinal specialist for further evaluation.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product was not returned to the manufacturing site; therefore, the complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no additional complaint was received from this production order.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|
Search Alerts/Recalls
|