Catalog Number S-60-060-120-P6
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2019001.The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other similar incidents.Based on the information provided, a definitive cause for the difficulty could not be determined.It may be possible that the introducer sheath was too close to the stent during deployment resulting in the stent partially deploying into the sheath; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified lesion in the left common femoral artery.During deployment of the 6.0x60 mm supera self expanding stent (ses), the stent jumped and deployed into the sheath.The stent was removed from the patient inside the sheath as it was not opposed to the vessel wall on the distal end.Another sheath and supera self expanding stent was used to complete the procedure.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA PERIPHERAL STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9451543
Report Number2024168-2019-14385
Device Sequence Number2159567
Product Code NIP
UDI-Device Identifier08717648211805
UDI-Public08717648211805
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Catalogue NumberS-60-060-120-P6
Device Lot Number9020561
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/19/2019
Initial Report FDA Received Date12/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
Patient Weight84
-
-