• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. FEMORAL/TIBIAL CHECKPOINT KIT -(STERILE); STEREOTAXIC DEVICE, ROBOTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. FEMORAL/TIBIAL CHECKPOINT KIT -(STERILE); STEREOTAXIC DEVICE, ROBOTICS Back to Search Results
Model Number 111645
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692)
Date of Event 11/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As part of the normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.Device not returned.
 
Event or Problem Description
Tibial and femoral checkpoints were left in patient after tka procedure.Case type: tka.Update: how many tibial checkpoints were left in the patient? 1.How many femoral checkpoints were left in the patient? 1.Did the checkpoints break and partial pieces were left in the patient? no, checkpoints did not break.Entire checkpoints were left in patient.
 
Additional Manufacturer Narrative
Reported event: tibial and femoral checkpoints were left in patient after tka procedure.Product evaluation and results: product inspection was not performed as product was not available for inspection.Product history review: product history review was not conducted as lot number was not provided.Complaint history review: complaint history review was not conducted as lot number was not provided.Conclusions: the alleged failure is due to the user error,since user has not followed the surgical protocol.There is no issue with the device.If device and/or additional information become available, this investigation will be reopened.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there have been no nc and capa associated with the product and failure mode reported in this event.
 
Event or Problem Description
Tibial and femoral checkpoints were left in patient after tka procedure.Case type: tka.Update: how many tibial checkpoints were left in the patient? 1.How many femoral checkpoints were left in the patient? 1.Did the checkpoints break and partial pieces were left in the patient? no, checkpoints did not break.Entire checkpoints were left in patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL/TIBIAL CHECKPOINT KIT -(STERILE)
Common Device Name
STEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key9451595
Report Number3005985723-2019-00897
Device Sequence Number2209472
Product Code OLO
UDI-Device Identifier00848486029494
UDI-Public00848486029494
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K143752
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number111645
Device Catalogue Number111645
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/12/2019
Supplement Date Received by Manufacturer01/27/2020
Initial Report FDA Received Date12/11/2019
Supplement Report FDA Received Date02/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-