• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US EXTERNAL ALIGN ANKLE CLAMP; KNEE INSTRUMENT : ALIGNMENT DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US EXTERNAL ALIGN ANKLE CLAMP; KNEE INSTRUMENT : ALIGNMENT DEVICES Back to Search Results
Model Number 2490-49-000
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported it was revealed at the time of washing the instruments postoperatively tka on (b)(6) 2019 that the tibial resection guide (p/n: 228701000) and the ankle clamp (p/n: 249049000) could not be separated because it could not be loosened the screw of the connection part.No further information is available.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy as not bsynthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTERNAL ALIGN ANKLE CLAMP
Common Device Name
KNEE INSTRUMENT : ALIGNMENT DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9451768
Report Number1818910-2019-122353
Device Sequence Number16795984
Product Code HXD
UDI-Device Identifier10603295125891
UDI-Public10603295125891
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2490-49-000
Device Catalogue Number249049000
Device Lot NumberSO2030232
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/19/2019
Supplement Date Received by Manufacturer01/29/2020
Initial Report FDA Received Date12/11/2019
Supplement Report FDA Received Date02/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-