| Model Number |
M00545890 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
Note: this report pertains to one of two hurricane rx dilation balloons used during the same procedure.It was reported to boston scientific corporation that the two hurricane rx dilation balloons were used in the pancreaticobiliary during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, it was noted that the two balloons both had a small hole, causing contrast leakage in the bile duct.The physician decided to not dilate further, and the procedure was completed with a plastic stent.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was confirmed to be stable.
|
| |
|
Event or Problem Description
|
|
Note: this report pertains to one of two hurricane rx dilation balloons used during the same procedure.It was reported to boston scientific corporation that the two hurricane rx dilation balloons were used in the pancreaticobiliary during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, it was noted that the two balloons both had a small hole, causing contrast leakage in the bile duct.The physician decided to not dilate further, and the procedure was completed with a plastic stent.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was confirmed to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Block a2: patient's exact age is unknown; however it was reported that the patient was over the age of 18.Block h6: problem code 1504 captures the reportable event of balloon pinhole.Block h10: investigation results a visual examination of the returned complaint device found that the balloon and the catheter did not have any visual defects and was in a good condition.It was noted that several quantity of contrast was in the balloon.Functional analysis was attempted to be performed; however, the balloon lumen was occluded and could not be unblocked.The balloon was then microscopically inspected and found that the balloon had a pinhole located at the proximal section of the body of the balloon.This failure is likely due to factors encountered during the procedure and/or the interaction with the scope that could have affected its performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
| |
|