MAUDE Adverse Event Report: COVIDIEN 16FR SALEM W/ GIENTRI PORT WIT; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 16FR SALEM W/ GIENTRI PORT WIT; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS Back to Search Results
Model Number 7771610E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/11/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event or Problem Description
The customer reported that the blue multi functional port was leaking the tube feed and medication out of the back during patient use.
 
Additional Manufacturer Narrative
H 3: evaluation summary: the device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported issue.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.The sample was received and after performing an evaluation according to procedure, the reported issue of leaking could not be duplicated in the sample.The reported condition could not be confirmed.The root cause for the reported condition cannot be specifically identified; therefore, corrective action will be limited to manufacturing awareness at this time.The current process is running according to product specifications, meeting quality acceptance criteria.The manufacturing plant will continue monitoring the process for any adverse trends that require immediate attention.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
16FR SALEM W/ GIENTRI PORT WIT
Common Device Name
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS
Manufacturer (Section D)
COVIDIEN
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
MDR Report Key9452050
Report Number9612030-2019-02375
Device Sequence Number2114387
Product Code PIF
UDI-Device Identifier10884521582736
UDI-Public10884521582736
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number7771610E
Device Catalogue Number7771610E
Device Lot Number1921802764
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 11/22/2019
Supplement Date Received by Manufacturer11/22/2019
Initial Report FDA Received Date12/11/2019
Supplement Report FDA Received Date06/29/2020
Patient Sequence Number1
Patient Weight67
-
-