| Model Number |
144110 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of the normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The surgeon began to manually ream the tibia for a universal baseplate with pressfit stem, when removing the tibial bone pins for the array assembly one of the bone pins broke inside the patient.Using x-ray the surgeon was able to locate the bone pin fixed inside cortical bone and was unable to retrieve it.Case type: tka.Surgical delay: 15 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: the surgeon began to manually ream the tibia for a universal baseplate with pressfit stem, when removing the tibial bone pins for the array assembly one of the bone pins broke inside the patient.Using x-ray the surgeon was able to locate the bone pin fixed inside cortical bone and was unable to retrieve it.Product evaluation and results: product inspection could not be performed as the product was not returned for evaluation.Product history review: review of the device history records for l/n w64539-1 indicate 969 devices were manufactured and accepted into final stock on 08/05/2019.No non-conformances were identified during inspection.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 144110, lot number w64539-1, shows no additional complaints related to the failure in this investigation.Conclusions: the failure could not be determined as the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there have been no nc and capa associated with the product and failure mode reported in this event.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
The surgeon began to manually ream the tibia for a universal baseplate with pressfit stem, when removing the tibial bone pins for the array assembly one of the bone pins broke inside the patient.Using x-ray the surgeon was able to locate the bone pin fixed inside cortical bone and was unable to retrieve it.Case type: tka.Surgical delay: = 15 minutes.
|
| |
|