The event unit was returned to applied medical for evaluation.Upon inspection, engineering confirmed that the lip of the bottle was damaged and the outer walls of the scope warmer bottle had collapsed.Based on the condition of the returned unit, it is likely that the outer walls of the scope warmer bottle had collapsed during the sterilization process due to pinholes in the scope warmer bottle.The pinholes were likely caused by repeated usage of the scope warmer bottle over time.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
|
|
Procedure performed: lap chole.At beginning of case, scope seal would not mount onto canister.Nurse said it was lip of canister was deformed and had been problematic for some time.As a result, during the case, the water became cold therefore device to serve purpose.Additional information received by email from applied medical representative on 25sep19: i am at the hospital now and have just spoken to [deputy theatre manager].She unfortunately cannot provide further information and the two theatre staff who were working that day are not here.Additional information received by phone from applied medical representative on 27sep19: you confirmed by phone that when you visited the hospital, they looked at the event record and did not find any information related to the patient status following the incident.In case there would have been any consequence on the patient, this would have been indicated in the hospital event record and communicated to you.Additional information received by email from applied medical representative on 17oct19: i have spoken to the staff member concerned, who confirms that the flask lip was damaged/bent before the procedure.This is at the beginning/setting up for the procedure as [company representative] has documented.Patient status: no indication of consequence on the patient.
|