| Model Number |
23295 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).Device evaluated by mfr.: a visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The returned device was attached to a boston scientific encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a pinhole located approximately 6mm distal of the distal markerband in the distal balloon sleeve.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 28-nov-2019.It was reported that balloon leak occurred.The target lesion was located in the severely tortuous and mildly calcified vein.A 3.0-4/4t/90 symmetry balloon catheter was advanced for dilation.However, the pressure did not increase during first inflation.Physician removed and checked the balloon, it was confirmed that a liquid leakage was seen in the tip side.The procedure was completed with another device.There were no patient complications reported.However, returned device analysis revealed balloon pinhole.
|
| |
|