| Model Number |
7133 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.00mm x 15mm emerge balloon catheter was selected for use.However, the device broke while introducing it in the guide catheter.There were no patient complications reported.No further information available.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.00mm x 15mm emerge balloon catheter was selected for use.However, the device broke while introducing it in the guide catheter.There were no patient complications reported.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.:returned product consisted of an emerge balloon catheter.The balloon was tightly folded.Microscopic examination revealed that the hypotube had numerous kinks.There were no separations identified.The reported separation was not returned.
|
| |
|