| Model Number |
G55736 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # k142688 investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
The needle kinked upon first puncture of the lesion.The needle was retracted back into the sheath, the doctor attempted a second puncture but failed because the needle was unable to come out from the sheath.Additional information confirmed that the needle was fully retracted with force prior to removal from patient.
|
| |
|
Event or Problem Description
|
|
The needle kinked upon first puncture of the lesion.The needle was retracted back into the sheath, the doctor attempted a second puncture but failed because the needle was unable to come out from the sheath.Additional information confirmed that the needle was fully retracted with force prior to removal from patient.File is fda reportable based on reporting precedence for this device family for the issue of ¿non-retraction of needle".
|
| |
|
Additional Manufacturer Narrative
|
|
Pma/510(k) #k142688, device evaluation: the echo-hd-22-c device of lot number c1542959 involved in this complaint was returned for evaluation, with the original packaging.The packaging was open on receipt.With the information provided, a physical examination and document based investigation was conducted.Lab evaluation the device related to this occurrence underwent a laboratory evaluation on the 27nov2019.There was a kink at the distal end of needle (approximately 3cm from the tip of the needle) document review prior to distribution, all echo-hd-22-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl a review of the relevant manufacturing records (c1542959) revealed no discrepancies that could have contributed to this complaint.Upon review of complaints, this failure mode has not occurred previously with this lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1542959.There is no evidence to suggest that the customer did not follow the instructions for use (ifu0077-4).Root cause review a definitive root cause could not be determined as the circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to the flexed position of the endoscope.Summary complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|