An elderly female patient with history of hypertension (htn), chest pain, recent traffic accident with t12 compression fracture and right (r) sided rib fractures presents to the hospital.While having a cardiac catheterization, the stent deployed in the left anterior descending (lad) artery.Unable to remove delivery balloon/ catheter.While attempting to remove, balloon tip fractured.Attempts made to snare and remove balloon.Pt coded.Balloon eventually retrieved.Unable to resuscitate patient expired.Procedure: cardiac cath."we were able to take the wire out but not the balloon which was stuck in the sever instant restenosis of the proximal left anterior (la) stent which is likely under expanded.I tried to use another coronary wire but could not pass by the balloon which was in the lad.We pulled the balloon hard and shaft broke with balloon still stuck in the lad stent.Patient had a pulseless electrical activity (pea) arrest and we took long sumariai wire and formed a snare through guide catheter to retrieve the stuck balloon and its shaft.There was no residual material were left in place afterwards.".
|