MEDINOL LTD. ELUNIR RIDAFOROLIMUS ELUTING CORONARY STENT SYSTEM; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 3.00 X 12
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Thrombosis (2100)
Date of Event 11/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the case detail report include information on four (4) sizes of the elunir product and since each mdr present only one (1) product, this report is related to one (1) out of four (4) elunir's products, involved in this event.Attempts to obtain additional information and identify the related product {out of the four (4) sizes of the elunir} will continue.Mfr report # for the other related cases: 3003084171-2019-00015, 3003084171-2019-00017, 3003084171-2019-00018.
 
Event or Problem Description
The initial case report was obtained on nov.17, 2019 and included the following "description as reported": "as reported by the sales rep, patient came in with a stemi on (b)(6) due to in stent thrombosis.Patient was on brilinta and went off due to financial reasons.After procedure the patient was put on plavix.It is unknown if the patient's dosage was correct for plavix or if the patients body didn't respond to the medication as well as it wasn't in the patients system long enough after the procedure which could have caused the elunir stents to shut down.Boston sci synergy stents were placed in the mid circ and mid lad" in addition, the information detailed in 'product information' in the same report included four (4) products.There have been several attempts to get information about which product within the reported four (4) was involved, but no answer has been received on this matter.Additional information received from distributor on dec 3, 2019 stated as follows: the device stored and handled as per the instruction for use (ifu).No damage was noted to the packaging of the device.The device prep as per the ifu 1 stent in mid circ and 1 stent in mid lad none were overlapped.The patient did not have any medical history found relevant for this case (e.G.Bleeding disorder, coronary artery disease.) neither site pre-dilated.Mid circ was inflated for 12atm for 11 seconds and mid lad was inflated for 14atm for 24 seconds.Good wall apposition was observed via angio.Patient is currently doing fine.Additional information received from distributor on dec 6, 2019 stated as follows: the stents for were implanted on 11/6 and it was 3.0x12 and 3.0x17 to the mid circ and also a 2.5x12 to the mid lad.
 
Event or Problem Description
Additional information (dated dec.08, 2019).The physician knew that the stemi was due to in-stent thrombosis based on the ekg from the stemi's and the fact the stents were all shut down in both patients.
 
Additional Manufacturer Narrative
Device history record (dhr) review performed on dec 23, 2019 have indicated the product was supplied meeting.Specifications.Ifu review (dated jan.15, 2020) no deviation/violation from ifu was reported.Final report (dated jan 16, 2020 ) overall conclusions: 1.The review of the dhr (device history record), along with the information from the customer indicate that the product was supplied meeting specifications.2.Stent thrombosis and myocardial infarction are well-known potential adverse events that may be associated with the implantation of a coronary stent in coronary arteries and are listed as such in section 8 of the elunir ifu.In this case the events may have been caused by the patient going off brilinta and/or by inappropriate dosage / insufficient time / absence of response with plavix.3.The current status of the patient is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUNIR RIDAFOROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
MEDINOL LTD.
beck tech bldg
8 hartom st.
jerusalem, 97775 08
IS  9777508
MDR Report Key9458205
Report Number3003084171-2019-00016
Device Sequence Number2146512
Product Code NIQ
UDI-Device Identifier07290107013703
UDI-Public07290107013703
Combination Product (Y/N)N
Initial Reporter CountryRP
PMA/510(K) Number
P170008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2020
Device Model Number3.00 X 12
Device Catalogue NumberLUN300R12US
Device Lot NumberLNRUS00171
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/17/2019
Supplement Date Received by Manufacturer12/23/2019
Initial Report FDA Received Date12/12/2019
Supplement Report FDA Received Date01/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age48 YR
Patient Weight79
-
-