• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DUR BANTAM 10D 22.225X40 LINER; DURALOC IMPLANT : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DUR BANTAM 10D 22.225X40 LINER; DURALOC IMPLANT : HIP POLY ACETABULAR LINERS Back to Search Results
Model Number 1241-40-526
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Cyst(s) (1800); Granuloma (1876); No Code Available (3191)
Date of Event 11/15/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event or Problem Description
Change of duraloc pe inlay right side.Doi: 2004, inlay was revised in 2012, date of latest inlay revision (b)(6) 2019 because of pe wear.A surgical delay was not reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. .
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: corrected:h6- no code available is to capture surgical intervention.Product complaint # (b)(4).Investigation summary: examination of the returned liner confirms the reported material wear.Nothing unusual or unexpected about the amount of wear is identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUR BANTAM 10D 22.225X40 LINER
Common Device Name
DURALOC IMPLANT : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9458373
Report Number1818910-2019-122705
Device Sequence Number11535860
Product Code LPH
UDI-Device Identifier10603295018926
UDI-Public10603295018926
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source foreign,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1241-40-526
Device Catalogue Number124140526
Device Lot Number113574
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2020
Initial Date Received by Manufacturer 11/21/2019
Supplement Date Received by Manufacturer01/03/2020
01/30/2020
Initial Report FDA Received Date12/12/2019
Supplement Report FDA Received Date01/14/2020
02/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient Weight55
-
-