• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS DE MEXICO SABER 6MM15CM 150; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS DE MEXICO SABER 6MM15CM 150; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 48006015X
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for analysis but has not yet been received.A device history record review was performed and showed that these lots of products met all requirements per the applicable manufacturing quality plan.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
During percutaneous transluminal angioplasty (pta) of a lesion in the superficial femoral artery (sfa), a the 6mm x 15cm x 150cm saber pta balloon catheter ruptured.There was no patient injury and the device will be returned for analysis.The device was opened in a sterile field.
 
Additional Manufacturer Narrative
During a percutaneous transluminal angioplasty (pta) of a lesion in the superficial femoral artery (sfa), a the 6mm x 15cm x 150cm saber pta balloon catheter ruptured.There was no patient injury.The device was opened in a sterile field.Additional procedural details were requested but unknown.The product was not returned for analysis.A product history record (phr) review of lot 82162595 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst¿ could not be confirmed as the product was not returned for analysis.The exact cause could not be conclusively determined.Vessel characteristics or procedural factors may have contributed to the reported event.However, with the limited amount of information provided and without the return of the device for analysis it is difficult to draw a clinical conclusion between the device and the event reported.According to the warnings in the safety information in the instructions for use ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABER 6MM15CM 150
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS DE MEXICO
circuito int nte 1820
juarez 32575
MX  32575
MDR Report Key9458615
Report Number9616099-2019-03399
Device Sequence Number3040202
Product Code LIT
UDI-Device Identifier20705032069622
UDI-Public20705032069622
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K971010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Model Number48006015X
Device Catalogue Number48006015X
Device Lot Number82162595
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 11/19/2019
Supplement Date Received by Manufacturer01/08/2020
Initial Report FDA Received Date12/12/2019
Supplement Report FDA Received Date01/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
-
-