Model Number ETLW1610C93E
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Injury (2348)
Date of Event 11/27/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Other relevant device(s) are: product id: e tlw1610c93e, serial/lot #: (b)(4), ubd: 14-aug-2021, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of an abdominal aortic ulcer of unknown size.It was reported that during the index procedure, the ipsilateral gate of the bifurcate was cannulated instead of the contralateral gate.As a result, both stent graft limbs were implanted in the same gate.The spin test (confirmed spinning the marking catheter) was performed and clearly visualized.The physician then implanted an endurant aui device to seal the bifurcate as treatment and completed a fem-fem bypass.As per the physician, the cause of the event was a procedural step error.No additional clinical sequelae were reported and the patient is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II ILIAC STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key9459047
Report Number9612164-2019-05147
Device Sequence Number11221142
Product Code MIH
UDI-Device Identifier00643169467606
UDI-Public00643169467606
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/14/2021
Device Model NumberETLW1610C93E
Device Catalogue NumberETLW1610C93E
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2019
Initial Report FDA Received Date12/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-