| Model Number |
1288210105 |
| Medical Device Problem Code |
Fogging (1253)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was fogging during a procedure, which could not be wiped away.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: blurry foggy image.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.Root cause: the most likely root causes for the issues found could be wear and tear and/or by the cable being pulled or bent hard during sterilization.Camera head is not producing a blurry foggy image.The cause of the blurry foggy image is the coupler.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was fogging during a procedure, which could not be wiped away.
|
| |
|