• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PRLNE BLU 4X24IN 6-0 D/A C-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PRLNE BLU 4X24IN 6-0 D/A C-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Catalog Number M8726
Medical Device Problem Codes Break (1069); Material Frayed (1262)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.How many devices had the issue (broken sutures and unraveled) in this procedure? please provide the specific issue for each device used in this single procedure? which product codes were noted be broken? which product codes were noted to be frayed? are actual samples or representative unopened samples of the devices for the reported lot numbers available for evaluation? if yes, please provide device return status and tracking information.Note: events reported via mw #2210968-2019-90847, 2210968-2019-90848.
 
Event or Problem Description
It was reported that the patient underwent cabg procedure on (b)(6) 2019 and suture was used to suture anastomosis.During the procedure, the suture may have broken and suture may have unraveled.A suture that unraveled was around the aorta and the surgeon spotted this before closing.The procedure was completed with no patient consequences reported.Additional information has been requested.
 
Additional Manufacturer Narrative
A manufacturing record evaluation was performed for the finished device batch and no non-conformances were identified.Additional h-3 summary: three unopened samples of product code m8726, lot peh424 were received for analysis.The complaint sample was not received for evaluation.The product code is multi-strand and required four strands double armed per packet.During the visual inspection, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, fraying suture or suture breakage were observed.A functional test was performed using and the tensile strength force was above the minimum requirement.Per the condition of the representative samples, no performance - breakage suture or fraying suture intra-op was found and the tested samples met the finished goods requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 4X24IN 6-0 D/A C-1
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key9461177
Report Number2210968-2019-90849
Device Sequence Number11539169
Product Code GAW
UDI-Device Identifier10705031045927
UDI-Public10705031045927
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Catalogue NumberM8726
Device Lot NumberPEH424
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2020
Initial Date Received by Manufacturer 11/21/2019
Supplement Date Received by Manufacturer01/06/2020
Initial Report FDA Received Date12/12/2019
Supplement Report FDA Received Date01/14/2020
Patient Sequence Number1
-
-