| Catalog Number |
0504880314 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the electrode was fractured during the procedure.The fracture did not take place inside the patient and no pieces fell into the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: the ending of electrode ring was fracture.Probable root cause: manufacturing/assembly error; severe shipping conditions; material/design error; constant irrigation flow is not maintained leading to limited field of view ; large anatomy; damaged or deformed o-ring; use error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that the electrode was fractured during the procedure.The fracture did not take place inside the patient and no pieces fell into the patient.
|
| |
|