| Model Number |
20188 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 6.0mmx18mmx150cm express sd renal/biliary stent was selected for use.However, before inserting the device into the patient's body, it was noticed that some stent struts appeared to be flared.The procedure was completed with a different device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an express sd stented catheter.The balloon was loosely folded.The stent was secured between the markerbands.The hypotube, outer shaft, inner shaft, balloon, stent and tip were visually and microscopically examined.The stent is damaged in the middle of the stent.Inspection of the device presented no damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 6.0mmx18mmx150cm express sd renal/biliary stent was selected for use.However, before inserting the device into the patient's body, it was noticed that some stent struts appeared to be flared.The procedure was completed with a different device.There were no patient complications reported.
|
| |
|