Model Number 7000D
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, a kink was identified 4 cm from the distal tip; which is distal to the transition point.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The most likely root cause is mishandling subsequent to distribution, including shipping/storage conditions or improper manipulation of the catheter.The instructions for use (ifu) state: do not introduce the tip folded into the guiding sheath.Do not exert excessive pressure during placement of catheter if unknown resistance is encountered.Do not attempt to use the reprocessed ep catheter prior to completely reading and understanding the directions for use.Avoid manual pre-bending of distal curve, as this may damage steering mechanism of steerable catheters.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported that the deflection on the electrophysiology catheter was broken upon insertion into the sheath.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
CATHETER, RECORDING, ELECTRODE, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS
1810 w. drake drive
tempe AZ 85283
MDR Report Key9462285
Report Number0002090040-2019-00034
Device Sequence Number13993034
Product Code NLH
UDI-Device Identifier00885825008697
UDI-Public00885825008697
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K052414
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/24/2021
Device Model Number7000D
Device Catalogue Number7000DRR
Device Lot Number3881208
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/15/2019
Initial Report FDA Received Date12/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Patient Age74 YR
-
-