Model Number N/A
Device Problem Separation Failure (2547)
Patient Problem No Information (3190)
Event Date 11/18/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical products: g7 pps ltd acet shell 50d , pn 010000662, ln 6549337.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2019-05526.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during an initial hip arthroplasty, a g7 cup inserter was stuck to the cup.This caused a 30 minute delay to wait for a new cup to complete the surgery attempts were made to obtain additional information; however, none was available.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned devices shows impact marks on the shaft and strike plate of the inserter along with several divots and dents but no obvious fracture on the tip of the thread.Material analysis confirmed the inserter material to be conforming.The shell showed use but no visible damage.Dhr was reviewed and no discrepancies were found.The root cause is unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Type of Device
HIP INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9465651
MDR Text Key189329126
Report Number0001825034-2019-05527
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 04/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110003450
Device Lot Number437740
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2019
Was the Report Sent to FDA? No
Date Manufacturer Received03/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-