| Catalog Number |
1012449-25 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Separation (1562)
|
| Health Effect - Clinical Codes |
Vessel Or Plaque, Device Embedded In (1204); Foreign Body In Patient (2687)
|
| Date of Event |
11/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number e2019001.The customer reported the device was discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a narrow lesion in the distal posterior tibial artery with moderate calcification.A 3.0 x 25 mm nc trek balloon dilation catheter (bdc) was advanced to the target lesion, and balloon angioplasty was performed.However, there was resistance during removal of the bdc, and excessive force was applied.The distal end of the catheter became detached inside the patient, and became embedded with the stent.The patient remains hospitalized and the separated portion was not retrieved.There was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during removal interaction with the moderately calcified and narrow anatomy resulted in the reported difficulty to remove.Manipulation of the device resulted in the reported material separation.As reported, an unspecified stent was deployed to embed the detached portion of the device.There is no indication of a product quality issue with respect to manufacture, design or labeling.H6: patient code 2687 removed.
|
| |
|
Event or Problem Description
|
|
Subsequent to the initially filed report, the following information was received: there was resistance during removal of the 3.0 x 25 mm nc trek balloon dilation catheter (bdc) due to the anatomy when the distal end of the bdc became detached.An unspecifed stent was deployed to embed the detached portion of the bdc.No additional information was provided.
|
| |
|