| Model Number |
MMT-723NAL |
| Medical Device Problem Codes |
Connection Problem (2900); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
12/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the reservoir lip ring was missing.The customer¿s blood glucose was 312 mg/dl at the time of incident.Customer stated the insulin pump had cosmetic damage.Customer reported that the reservoir was not able to lock in place.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan as per health care professional instruction.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with cracked reservoir tube lip noted.Device passed displacement test.The test reservoir was able to lock in place.
|
| |
|