MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723NAL
Medical Device Problem Codes Connection Problem (2900); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 12/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the reservoir lip ring was missing.The customer¿s blood glucose was 312 mg/dl at the time of incident.Customer stated the insulin pump had cosmetic damage.Customer reported that the reservoir was not able to lock in place.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan as per health care professional instruction.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device received with cracked reservoir tube lip noted.Device passed displacement test.The test reservoir was able to lock in place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9466092
Report Number2032227-2019-132395
Device Sequence Number4115078
Product Code OYC
UDI-Device Identifier00643169077713
UDI-Public(01)00643169077713
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723NAL
Device Catalogue NumberMMT-723NAL
Device Lot NumberA6723NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2020
Initial Date Received by Manufacturer 12/10/2019
Supplement Date Received by Manufacturer03/11/2020
Initial Report FDA Received Date12/13/2019
Supplement Report FDA Received Date03/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age54 YR
Patient Weight68
-
-